A clip-on extensometer is still the most cost-effective way to measure strain accurately on a universal testing machine. It attaches directly to the specimen, measures the change in gauge length with a strain gauge or LVDT transducer, and feeds the testing software so modulus, yield and elongation are calculated correctly. This guide explains how to choose one and what information is needed to configure it for your machine.

Why measure strain with an extensometer?

Crosshead movement includes the stretch of the load frame, load cell, grips and the specimen shoulders, not only the gauge section. That makes strain based on crosshead travel too high and modulus far too low. An extensometer measures only the gauge section, which is why standards such as ASTM D638, ASTM E8 and ASTM D3039 require one for modulus and yield.

Clip-on or video?

Clip-on vs video extensometers
Clip-on extensometerVideo extensometer
Typical priceAbout $4,000 to $5,000 per unitSystems start around $20,000
ContactKnife edges on the specimenNon-contact
Gauge lengthsFixed per model, some adjustableSet in software
Best forMetals, rigid plastics, composites, one main standardSoft, thin or high-elongation materials, many standards, high throughput
Through fractureUsually removed before break on brittle or violent failuresStays on to fracture

If you test several gauge lengths, soft or thin specimens on which the weight of a clip-on affects the result, or run high volumes with many operators, a video extensometer such as the ONE M1 or a 3D extensometer may be the better long-term investment.

Our clip-on extensometers

We are an authorized distributor of Epsilon extensometers, made in the USA, and configure each unit for the customer’s machine.

Video: Automatic balanced extensometer demonstration on plastics

What we need to configure your extensometer

  1. Material, specimen and standard, for example ASTM D638 Type I plastic, an ASTM E8 round or flat specimen, or an ASTM D3039 composite strip. This sets the arm geometry and the specimen thickness or diameter the extensometer must fit.
  2. Gauge length. Common choices are 25 mm, 50 mm, 1 inch and 100 mm. It must match the standard and leave room between the grips.
  3. Maximum strain. Roughly 10% for many metals, around 50% for semi-rigid plastics and 100% or more for elastomers. This sets the measuring range.
  4. Your testing machine. The model number and a photo of the extensometer input on the back of the controller, so we can match the connector, excitation and calibration. We configure units for Instron, MTS, ZwickRoell, Tinius Olsen, Galdabini and other machines.

Example configuration

Example: clip-on extensometer for ASTM D638
ItemCustomer requirement
ApplicationASTM D638 tensile testing of polypropylene
SpecimenType I, 165 mm × 13 mm × 3 mm
Gauge length50 mm
StrainUp to 100% elongation
MachineInstron 3367
Connector3-pin round DIN, 5 V excitation

We supplied an Epsilon 3542 with a 50 mm gauge length and the matching connector, pre-calibrated so the customer could test immediately on delivery.

Accuracy classes

ASTM E83 and ISO 9513 classify extensometers by accuracy. Metals testing to ASTM E8 needs at least Class B-2 for yield strength; plastics modulus to ISO 527 calls for Class 0.5 or better at small strains. Ask for a calibration certificate to the class your standard requires, and recalibrate at the interval your quality system sets.

Tips for good results

  • Attach the extensometer at the center of the gauge section, square to the specimen.
  • Zero it after attaching and before the test starts.
  • Remove it before fracture on brittle or violent breaks unless it is rated to stay on.
  • Store it in its case and check knife edges for wear.

Frequently asked questions

How much does a clip-on extensometer cost?

A configured clip-on extensometer typically costs about $4,000 to $5,000. Video extensometer systems start around $20,000.

Will a clip-on extensometer work with my Instron, MTS or ZwickRoell machine?

Yes, when it is configured with the right connector, excitation and calibration for your controller. Send us the model number and a photo of the extensometer input.

What gauge length do I need?

The one your test standard specifies: 50 mm for ASTM D638 Type I and ISO 527 multipurpose specimens, 50 mm or 2 inches for ASTM E8 sheet specimens, for example.

Can a clip-on extensometer measure to break?

Many can on ductile materials, but on brittle or violent failures it is safer to remove it after yield. A video or 3D extensometer can stay on through fracture.

What accuracy class do I need?

ASTM E8 requires at least Class B-2 of ASTM E83 for yield strength. Plastics modulus to ISO 527 needs Class 0.5 or better. Check your standard.

Get a quote

Interested in an immediate price quote or an application discussion? Email christhomas@universalgripco.com or frank@universalgripco.com, or call or text us at 978-882-1480. Send us your material and standard, gauge length, maximum strain, machine model and a photo of the extensometer input, and we will quote a configured extensometer.